SYMPOSIA PAPER
Published:
01 January 2010
STP48992S
Fatigue to Fracture: An Informative, Fast, and Reliable Approach for Assessing Medical Implant Durability
SourceThis paper compares the differences between the “test to success” and “test to fracture” paradigms for medical implant durability assessment. It recognizes that test to success can be one of the tests within test to fracture, in addition to a series of shorter cycle, higher force/deformation tests. In addition, by “testing to fracture,” one can validate the fatigue analysis capability, learn the consequences of in vitro fractures, and provide information that can either be used for product improvement or for clinicians to make life-saving decisions.
Author Information
Gong, X., -Y.
Medical Implant Mechanics, LLC, Aliso Viejo, CA
Chwirut, D., J.
CardioMed Device Consultants, Rockville, MD
Mitchell, M., R.
Mechanics and Materials Consulting, LLC, Flagstaff, AZ
Choules, B., D.
Med Institute, West Lafayette, IN
Related
Reprints and Permissions
Details
Developed by Committee: E08
Pages: 108–121
DOI: 10.1520/STP48992S
ISBN-EB: 978-0-8031-8418-3
ISBN-13: 978-0-8031-7501-3