SYMPOSIA PAPER
Published:
01 January 2007
STP45471S
Spent Fuel Monitoring with Silicon Carbide Semiconductor Neutron/Gamma Detectors
SourceSilicon carbide semiconductor radiation detectors have been demonstrated for neutron and gamma-ray monitoring of spent nuclear fuel. Neutrons and gamma rays were monitored simultaneously over a 2050-h period, resulting in a gamma dose of over 6000 Gy to the SiC detector. No changes in the neutron and gamma-ray sensitivity were observed as a result of the gamma-ray exposure. After the spentfuel measurements, the absolute neutron sensitivity was determined in a standard neutron field, and the degree of gamma-ray and neutron spectrum overlap were determined through exposures in intense gamma ray fields. No overlap was observed for gamma-ray dose rates up to 100 G/h.
Author Information
Natsume, T.
Mitsubishi Heavy Industries, LTD., Hyogo-ken, Japan
Doi, H.
Nuclear Development Corporation 622-12, Tokaimura Ibaraki, Japan
Ruddy, F., H.
Westinghouse Electric Company, Pittsburgh, PA
Seidel, J., G.
Westinghouse Electric Company, Pittsburgh, PA
Dulloo, A., R.
Westinghouse Electric Company, Pittsburgh, PA
Related
Reprints and Permissions
Details
Developed by Committee: E10
Pages: 424–431
DOI: 10.1520/STP45471S
ISBN-EB: 978-0-8031-6242-6
ISBN-13: 978-0-8031-3412-6