Sterility and Pyrogenicity Issues of Synthetic Vascular Grafts
SourceCommon sterilization techniques, validation methods, and the physical and biologic effects of sterilization on synthetic vascular graft materials are reviewed. Pyrogenicity issues are discussed. A series of endotoxin-contaminated polyester externally supported grafts were tested in a canine model to assess the clinical significance of pyrogen contamination. Systemic pyrogenicity was observed, but gross thrombus formation was unaffected. The pathogenesis of graft infections are then reviewed.
Author Information
Helmus, MN
Arthur D. Little, Inc., Cambridge, MA
Botan, EA
Bard Cardiosurgery Division, C. R. Bard, Inc., Billerica, MA
Malone, JM
Tucson Veteran's Administration Medical Center, Tucson, AZ
Botzko, KM
, Reading, MA
Reinert, RL
Tucson Veteran's Administration Medical Center, Tucson, AZ
Bevans-Lynch, R
Bard Cardiosurgery Division, C. R. Bard, Inc., Billerica, MA
Brendel, K
College of Medicine, University of Arizona, Tucson, AZ
Duhamel, RC
College of Medicine, University of Arizona, Tucson, AZ
Snyder, RW
Advanced Vascular Technologies, Inc., Webster, TX
Related
Reprints and Permissions
Details
Developed by Committee: F04
Pages: 236–249
DOI: 10.1520/STP33295S
ISBN-EB: 978-0-8031-4966-3
ISBN-13: 978-0-8031-0462-4