The purpose of this study was to characterize the prevalence of taper damage in retrieved modular shoulder arthroplasty components. We matched this cohort with historical and contemporary total hip replacements. The components were matched primarily for implantation time, but also for alloy combination (mixed versus similar) and patient age (as a proxy for patient activity), when available. Modular components were evaluated for mechanically assisted crevice corrosion using a semiquantitative four-point scoring system. In total, 92 anatomical humeral heads and 67 associated stems with an average implantation time of 4.6 ± 4.3 years were analyzed. Additionally, 79 femoral heads and 61 associated stems with an average implantation time of 4.5 ± 4.4 years were matched to shoulder components and analyzed. Design and patient factors were assessed as potential predictors of taper damage. Evidence of damage was observed on 46 % of humeral heads and 66 % of the humeral stems. In the hips, 47 % of femoral heads and 44 % of femoral stems had evidence of damage. Mixed alloy pairs (p < 0.0001; stem side only), taper rigidity (p < 0.0001), and implantation time (p = 0.01) were associated with taper corrosion. The results from this study supported the prevalence of taper damage that is similar between anatomical shoulder and conventional hip arthroplasty. The clinical implications of these findings have not yet been determined.
Author Information
Day, J., S.
Drexel University, Implant Research Center, Philadelphia, PA, US
MacDonald, D., W.
Drexel University, Implant Research Center, Philadelphia, PA, US
Abboud, J., A.
Rothman Institute at Jefferson, Philadelphia, PA, US
Williams, G., R.
Rothman Institute at Jefferson, Philadelphia, PA, US
Rimnac, C., M.
Case Western Reserve University, Center for the Evaluation of Implant Performance, Depts. of Mechanical and Aerospace Engineering and Orthopaedics, Cleveland, OH, US
Kraay, M., J.
University Hospitals Case Medical Center, Cleveland, OH, US
McCloskey, R., C.
The Catholic University of America, Washington, DC, US
Arnholt, C., M.
Drexel University, Implant Research Center, Philadelphia, PA, US
Kurtz, S., M.
Drexel University, Implant Research Center, Philadelphia, PA, US
Domestic orders are delivered via United Parcel Service (UPS) or United States Postal Service (USPS). Transit
times average 3 to 5 business days. Please be aware that UPS will not deliver packages to Post Office Boxes.
International orders are delivered via courier post services which can be either a postal service, courier
service, or a combination of both. Standard Service is untraceable. Please allow 4-7 weeks for delivery.
Please be aware that carriers will not deliver packages to Post Office Boxes. Because of the variability of
customs processes and procedures in different countries, ASTM International cannot guarantee transit times to
international destinations. Customs duty and taxes are the responsibility of the consignee.
Shipping & Handling charges follow the rate schedule, below:
Order Total
Shipping & Handling Fee (US Domestic)
Up to $50.00
$18.72
$50.01 to $100.00
$20.80
$100.01 to $150.00
$29.52
$150.01 to $250.00
$39.09
$250.01 to $500.00
$56.25
$500.01 to $750.00
$76.42
$750.01 to $1000.00
$93.15
$1000.01 to $1500.00
$121.27
$1500.01 to $2500.00
$158.38
$2500.01 to $4999.00
$209.04
$5000.00 to higher
FREE
Order Total
Shipping & Handling Fee (International)
Up to $50.00
$68.72
$50.01 to $100.00
$70.80
$100.01 to $150.00
$79.52
$150.01 to $250.00
$89.09
$250.01 to $500.00
$106.25
$500.01 to $750.00
$126.42
$750.01 to $1000.00
$143.15
$1000.01 to $1500.00
$171.27
$1500.01 to $2500.00
$208.38
$2500.01 to $4999.00
$259.04
$5000.00 to higher
FREE
Shipping and Handling charges are approximate. Additional charges may be incurred if your order requires multiple shipments. This does not apply to complete sets and sections.