SYMPOSIA PAPER
Published:
01 January 1994
STP15137S
Reusability of CaF :Mn Thermoluminescence Dosimeters for Photon Irradiations at High Absorbed-Dose Levels
SourceThe effects of high absorbed dose on the reusability of CaF2:Mn thermoluminescence dosimeters (TLDs) are investigated to determine a recommended upper limit on absorbed dose for TLDs that are to be reused. This investigation examines degradation in the uniformity of response and changes in sensitivity of a batch of TLDs when exposed to gamma-radiation doses ranging from 1 Gy to 1000 Gy, and confirms earlier work suggesting that CaF2:Mn TLDs should not be reused in applications where cumulative absorbed doses are likely to exceed 100 Gy.
Author Information
Vehar, DW
Radiation Metrology Laboratory, Sandia National Laboratories, Albuquerque, NM
Related
Reprints and Permissions
Details
Developed by Committee: E10
Pages: 433–440
DOI: 10.1520/STP15137S
ISBN-EB: 978-0-8031-5281-6
ISBN-13: 978-0-8031-1899-7