Standard Active Last Updated: Oct 01, 2025 Track Document
ASTM F2009-25

Standard Test Method for Determining the Axial Disassembly Force of Taper Connections of Modular Prostheses

Standard Test Method for Determining the Axial Disassembly Force of Taper Connections of Modular Prostheses F2009-25 ASTM|F2009-25|en-US Standard Test Method for Determining the Axial Disassembly Force of Taper Connections of Modular Prostheses Standard new BOS Vol. 13.01 Committee F04
$ 72.00 In stock
ASTM International

Significance and Use

4.1 This test method helps to assess the axial locking force of a modular taper. Some types of devices that can utilize this type of connection are the modular shoulder and modular hip prostheses. Additional means of evaluating the locking mechanisms of tapers can be appropriate, depending upon the design of the device.

4.2 This test method may not be appropriate for all implant applications. The user is cautioned to consider the appropriateness of the test method in view of the materials and design being tested and their potential application.

4.3 While this test method can be used to measure the force required to disengage tapers, any comparison of such data for various component designs must take into consideration the size of the implant and the type of locking mechanism evaluated.

Scope

1.1 This test method establishes a standard methodology for determining the force required, under laboratory conditions, to disassemble tapers of implants that are otherwise not intended to release. Some examples are the femoral components of a total or partial hip joint replacement or humeral components of an anatomic total or partial shoulder joint replacement in which the head and base component are secured together by a self-locking taper.

1.2 In this test method, it is recognized that actual loading during surgery and in vivo is quite varied, and that no one set of experimental conditions can encompass all possible variations. Thus, the test method included here represents a simplified model for purposes of comparison between designs and materials. Test methods involving more complex systems may require additional justification.

1.3 This test method has been developed primarily for evaluation of metal and ceramic head designs on metal tapers but can have application to other materials and designs.

1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.

1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

Price:
Contact Sales
Related
Reprints and Permissions
Reprints and copyright permissions can be requested through the
Copyright Clearance Center
Details
Book of Standards Volume: 13.01
Developed by Subcommittee: F04.22
Pages: 6
DOI: 10.1520/F2009-25